EKG Paper, Calibration and Heart-Rate Methods

At 25 mm/s and 10 mm/mV a small box is 0.04 s and 0.1 mV and a large box 0.20 s and 0.5 mV. Check the speed and the 10 mm calibration pulse first, then use 1500 or 300 for a regular rhythm and the 6-second count for an irregular one.

The grid at standard settings (25 mm/s, 10 mm/mV)

unittimeheight
Small box (1 mm)0.04 s0.1 mV
Large box (5 mm)0.20 s0.5 mV
5 large boxes1 s
30 large boxes6 s (a 6-second strip)
Calibration pulse10 mm tall = 1 mV

Horizontal is time; vertical is voltage. Measure height from the baseline.

Preview only. The full sheet also covers: Non-standard settings; Calibration pulse check; Rate methods; Count-off (300) values; 1500 values; Filters (AHA/ACC/HRS recommendations).