EKG Acquisition practice questions for the CET exam

EKG Acquisition is 44 of the 100 scored questions (44%) on the CET exam. Try 10 free questions below: tap an answer to check it, then read why the other options are wrong and the sources.

10 EKG Acquisition questions with answers

10 questions. Tap an answer to check it, or open the explanation under each one.

Question 1

A patient's resting systolic pressure was 140 mm Hg, and it rose to 152 mm Hg in stage 2. In stage 3 it reads 126 mm Hg, and he says his chest feels tight. What should the technician do?

  1. Speed up the treadmill to raise the pressure
  2. Alert the supervising provider immediately
  3. Note it and recheck at the end of the next stage
  4. Tell him the drop is normal for this stage
Show answer and explanation

Correct answer: B. Alert the supervising provider immediately

Systolic pressure 14 mm Hg below the resting value while the workload rises, together with chest tightness suggesting ischemia, meets an absolute indication to stop (a fall of more than 10 mm Hg from baseline with other evidence of ischemia); the provider is alerted at once.

Why the other options are wrong

  • A. Speed up the treadmill to raise the pressure: More workload is dangerous when pressure is falling with symptoms.
  • C. Note it and recheck at the end of the next stage: Waiting another stage delays action on an absolute stop criterion.
  • D. Tell him the drop is normal for this stage: A falling systolic pressure during exercise is not a normal response.

Question 2

What instruction does the technician give just before pressing the button to acquire a resting EKG?

  1. Take deep, quick breaths
  2. Lift your head to watch the screen
  3. Clasp your hands on your chest
  4. Lie still, relax and stay quiet
Show answer and explanation

Correct answer: D. Lie still, relax and stay quiet

Any movement can alter the tracing, so the patient lies still, relaxed and quiet while the tracing is recorded.

Why the other options are wrong

  • A. Take deep, quick breaths: Rapid deep breathing moves the chest and baseline.
  • B. Lift your head to watch the screen: Lifting the head tenses the neck and chest muscles.
  • C. Clasp your hands on your chest: Clasped hands tense the arm muscles and can lie on chest electrodes.

Sources

Question 3

Studies cited by AACN found that 89% to 99% of ECG monitor alarms were false or clinically insignificant. What is the safety risk of this for a telemetry unit?

  1. Monitors wear out sooner from frequent alarms
  2. Patients cannot sleep, so their rhythm becomes irregular
  3. Staff become desensitized and may miss a real alarm
  4. Electrodes dry out faster when alarms sound often
Show answer and explanation

Correct answer: C. Staff become desensitized and may miss a real alarm

Alarm fatigue is a patient safety risk: frequent false or insignificant alarms desensitize staff, so a true alarm may be delayed or missed.

Why the other options are wrong

  • A. Monitors wear out sooner from frequent alarms: Device wear is not the safety problem described.
  • B. Patients cannot sleep, so their rhythm becomes irregular: The described risk concerns staff response, not the patient's rhythm.
  • D. Electrodes dry out faster when alarms sound often: Alarm frequency does not affect electrode gel.

Question 4

Before pressing record, the technician watches the live tracing on the machine's screen for about 10 seconds. What is this check for?

  1. To let the machine warm up the thermal paper first
  2. To measure the patient's blood pressure
  3. To confirm good contact and catch error messages
  4. To give the gel time to dry out
Show answer and explanation

Correct answer: C. To confirm good contact and catch error messages

After connecting the leads, observe the displayed tracing to make sure electrode contact is good, and watch for any error messages before reviewing the recording.

Why the other options are wrong

  • A. To let the machine warm up the thermal paper first: Watching the screen does not prepare the paper.
  • B. To measure the patient's blood pressure: An electrocardiograph does not measure blood pressure.
  • D. To give the gel time to dry out: Dry gel worsens contact; the check is for a good signal.

Sources

Question 5

Which electrode is the positive pole for lead aVF?

  1. Right leg
  2. Left arm
  3. Right arm
  4. Left leg
Show answer and explanation

Correct answer: D. Left leg

aVF (augmented voltage, foot) uses the left leg electrode as its positive pole; aVR uses the right arm and aVL the left arm.

Why the other options are wrong

  • A. Right leg: The right leg is the reference electrode.
  • B. Left arm: The left arm is positive for aVL.
  • C. Right arm: The right arm is positive for aVR.

Question 6

When should the patient be taught how to use the effort rating scale for a treadmill test?

  1. During the recovery period
  2. At the first sign of chest discomfort
  3. Before the exercise begins
  4. Only if the patient asks about it
Show answer and explanation

Correct answer: C. Before the exercise begins

Before an exercise test the patient should be familiarized with the symptom and effort rating scales so they can report effort accurately during exercise.

Why the other options are wrong

  • A. During the recovery period: Ratings are needed during exercise, so teaching after is too late.
  • B. At the first sign of chest discomfort: Teaching a scale during a symptom wastes time and gives unreliable ratings.
  • D. Only if the patient asks about it: Familiarization is a routine step for every patient.

Question 7

A 12-lead prints without any calibration pulse. What should the technician do?

  1. Rerun it so the pulse prints
  2. Write 'standard gain' in the header
  3. Accept it, since 10 mm/mV is the default
  4. Draw a 10 mm mark by hand at the start
Show answer and explanation

Correct answer: A. Rerun it so the pulse prints

The calibration standard should be recorded on each resting ECG; without it, amplitudes cannot be checked, and research reading centers treat a missing mark as a technical defect.

Why the other options are wrong

  • B. Write 'standard gain' in the header: A note does not replace the recorded calibration.
  • C. Accept it, since 10 mm/mV is the default: The default may have been changed; only the printed pulse proves the setting.
  • D. Draw a 10 mm mark by hand at the start: A hand-drawn mark does not show what the machine actually recorded.

Sources

Question 8

Which item in an exam room is a listed source of electrical interference on an EKG?

  1. A cotton blanket
  2. A plastic clipboard
  3. A wooden step stool
  4. A fluorescent lamp
Show answer and explanation

Correct answer: D. A fluorescent lamp

Listed sources include cell phones near the module, electric beds, surgical and fluorescent lamps, and other electrical equipment in the room.

Why the other options are wrong

  • A. A cotton blanket: A blanket has no electrical supply.
  • B. A plastic clipboard: A plastic clipboard produces no electrical field.
  • C. A wooden step stool: A wooden stool has no electrical supply.

Question 9

The technician is connecting all ten lead wires. Which habit best prevents clipping a wire onto the wrong electrode?

  1. Using one fixed connection order and reading each wire's label
  2. Connecting the wires that reach most easily first
  3. Letting the machine detect which wire is on which electrode
  4. Matching wires by colour alone
Show answer and explanation

Correct answer: A. Using one fixed connection order and reading each wire's label

A uniform attachment sequence lowers the risk of lead reversal, and each connection is then checked one by one against its electrode.

Why the other options are wrong

  • B. Connecting the wires that reach most easily first: Convenience order invites mix-ups.
  • C. Letting the machine detect which wire is on which electrode: The machine does not sort wires; a reversal is recorded as is.
  • D. Matching wires by colour alone: Colour codes differ between systems, so colour alone can mislead.

Question 10

A cardiology lab keeps finding V1 and V2 placed too high on its tracings. What do the AHA/ACCF/HRS recommendations advise?

  1. Buying a new electrocardiograph
  2. Periodic retraining of staff in electrode and patient positioning
  3. Letting the computer correct the placement
  4. Moving V1 and V2 to the 3rd space for everyone
Show answer and explanation

Correct answer: B. Periodic retraining of staff in electrode and patient positioning

The recommendations call for periodic retraining of personnel in skin preparation, electrode positioning and patient positioning to prevent precordial misplacement and lead switches.

Why the other options are wrong

  • A. Buying a new electrocardiograph: The machine is not the cause of misplacement.
  • C. Letting the computer correct the placement: Software cannot move electrodes.
  • D. Moving V1 and V2 to the 3rd space for everyone: The 3rd space is still too high.

EKG Acquisition terms and cheat sheets

More EKG practice

Take the free 25-question EKG practice test or the free 100-question full-length test, or see what to study first in the EKG study guide.